Automated Pressure Calibrators
Additel 761
Pressure Ranges
Electrical Measurement and Source Accuracy
SPECIFICATIONS
* One year accuracy (including yearly stability) except ADT761-LLP which is 0.05%FS calibration accuracy and 0.05%FS yearly stability.
* * Additel 761 calibrators support 160 series intelligent digital pressure modules that are available for gauge, vacuum and absolute pressure
from -15 psi to 10,000 psi (-1 bar to 700 bar). Accuracy from 0.025%FS includes operation over 14°F to 122°F (-10°C to 50°C), one year
stability and calibration uncertainty. For detailed specication refer to 160 series pressure modules datasheet.
Type Model Pressure Range
Pressure Module 1 Pressure Module 2
Range Accuracy Range Accuracy
Low/Differential
Pressure
ADT761-LLP
-25 to 25 mbar
(-10 to 10 inH
2
O)
-2.5 to 2.5 mbar
(-1 to 1 inH
2
O)
0.05%FS
-25 to 25 mbar
(-10 to 10 inH
2
O)
0.05%FS
ADT761-D
-0.95 to 1 bar
(-13.5 to 15 psi)
-25 to 25 mbar
(-10 to 10 inH
2
O)
0.05%FS
-0.95 to 1 bar
(-13.5 to 15 psi)
0.02%FS
Gauge Pressure
ADT761-L
-0.95 to 7 bar
(-13.5 to 100 psi)
-0.95 to 2.5 bar
(-13.5 to 35 psi)
0.02%FS
0 to 7 bar
(0 to 100 psi)
0.02%FS
ADT761-M
-0.90 to 25 bar
(-13 to 375 psi)
-0.9 to 2.5 bar
(-13 to 35 psi)
0.02%FS
0 to 25 bar
(0 to 375 psi)
0.02%FS
ADT761-H
-0.90 to 40 bar
(-13 to 600 psi)
-0.9 to 2.5 bar
(-13 to 35 psi)
0.02%FS
0 to 40 bar
(0 to 600 psi)
0.02%FS
Gauge/Absolute
Pressure
ADT761-LA
-0.95 to 7 bar
(-13.5 to 100 psi)
-0.95 to 2.5 bar
(-13.5 to 35 psi)
0.02%FS
-0.95 to 7 bar
(-13.5 to 100 psi)
0.02%FS
0.05 to 8 bar.a
(1.2 to 115 psi.a)
0.05 to 3.5 bar.a
(1.2 to 50 psi.a)
0.03%FS
0.05 to 8 bar.a
(1.2 to 115 psi.a)
0.025%FS
ADT761-MA
-0.90 to 25 bar
(-13 to 375 psi)
-0.9 to 2.5 bar
(-13 to 35 psi)
0.02%FS
-0.9 to 25 bar
(-13 to 375 psi)
0.02%FS
0.1 to 26 bar.a
(1.7 to 390 psi.a)
0.1 to 3.5 bar.a
(1.7 to 50 psi.a)
0.03%FS
0.1 to 26 bar.a
(1.7 to 390 psi.a)
0.025%FS
ADT761-HA
-0.90 to 40 bar
(-13 to 600 psi)
-0.9 to 2.5 bar
(-13 to 35 psi)
0.02%FS
-0.9 to 40 bar
(-13 to 600 psi)
0.02%FS
0.1 to 41 bar.a
(1.7 to 615 psi.a)
0.1 to 3.5 bar.a
(1.7 to 50 psi.a)
0.03%FS
0.1 to 41 bar.a
(1.7 to 615 psi.a)
0.025%FS
Barometric pressure ADT761-BP 100 to 1,200 hPa 100 to 1,200 hPa 0.01%FS NA NA
Range Resolution Accuracy
Voltage Measurement ±30.0000V 0.1mV ±(0.01%RD + 1.5mV)
Current Measurement ±30.0000mA 0.1μA ±(0.01%RD + 1.5μA)
Current Source 0.000 to 22.000mA 1μA ±(0.02%RD + 2.2μA)
Switch Test If the switch has detective Voltage, the range is from 3V to 24V
DC 24V output 24V ± 0.5V, max: 50mA